Modes
The trace SN cube is optimized for the analysis of:- SN
- S
- N
Analytical specifications
Sample volume
Injection volume free adjustable from 3-80 µl
Analysis time
~ 4-6 min per parameter, self-optimizing according to element content and sample weight. Depending on sample type, analysis mode, and configuration.
Element concentration range
Large measuring range:
S: up to 1,000 ppm, 6 ppb LOD
N: up to 1,000 ppm, 15 ppb LOD
Precision
< 2 % rel. for 10 ppm standard (S and N)
Technical specifications
The trace SN cube offers high-temperature combustion in the carrier gas at up to 1,150 °C, with direct injection of liquid samples into separate ports for N and S determination.
Purge and trap technique
This option allows users to enable the determination of very low S concentrations, even when N levels are high.
Sample introduction
Flexible sample introduction:
- Manually with a syringe
- Integrated 50* positions autosampler (not temperature controllable) for liquids
- Integrated 56* positions autosampler (temperature controllable) for liquids
- Integrated 80* or 120* positions autosampler for solids
Required gases
Synthetic air, argon, oxygen for N
Synthetic air for S
Detector
High sensitive UV-F and CL detector, linear over entire range
Maximum furnace temperature
1,200 °C
Instrument control
Windows® based trace SN cube operating software with LIMS integration and auto sleep and wake-up function for automated and unattended overnight operation. 21 CFR Part 11 functionality available.*
Dimensions
47 × 55 × 57 (W × D × H) without detectors, autosampler
* optionally available
Accessories
Electronic balance
Balance and elemental analyzer form a system. We offer analytical balances from the leading manufacturers with our instruments. Please ask us for further information.
SN standards dissolved in toluene
We offer a variety of SN standards dissolved toluene with concentrations from 1 µg/ml up to 1,000 µg/ml.