Modes
The rapid CS cube is optimized for the analysis of:- CS
- C
- S
- Cl*
- TIC in solids*
* optionally available
Analytical specifications
Weighing range
Large weighing range:
The instrument’s range encompasses micro (< 1 mg) and macro (1 g) applications, depending on the sample type.
Precision
< 1 % relative (homogeneous substance), depending on sample type, analysis mode, and configuration
Element concentration range
Large dynamic concentration range:
C: up to 40 mg absolute, from 0 - 100 %
S: up to 20 mg absolute, from 0 - 100 %
Cl*: up to 1.2 mg, from 0 - 100 %
Analysis time
~ 3-5 min, self-optimizing according to element content and sample weight. Depending on sample type, analysis mode, and configuration.
* optionally available
Technical specifications
Quantitative high-temperature decomposition
The rapid CS cube functions at a permanent temperature near 1,200 °C, reaching 1,800 °C during combustion due to the combustion enthalpy of the tin capsules. In pure oxygen, the contained sulfur is fully converted into SO2 with 100 % recovery, even in the case of BaSO4. Carbon is converted to CO2 and detected by a real non-dispersive IR detector.
Sample introduction
- Integrated 60, 80*, or 120* positions autosampler (for solids and liquids in capsules), reloadable during analysis
- Integrated 2 ml vial, 50 positions liquid autosampler with rinse and waste vial*, reloadable during analysis
Instrument control
Windows® based rapid CS cube operating software with LIMS integration and auto sleep and wake-up function for automated and unattended overnight operation. 21 CFR Part 11 functionality available.*
Detector
- SO2 and CO2 specific infrared detector for sulfur and carbon analysis
- Electrochemical detector* for chlorine
Maximum furnace temperature
1,200 °C (whereby 1,800 °C is briefly reached at the combustion point when using tin foil)
Required gases
Oxygen or synthetic air
Dimensions
48 x 55 x 55 cm (W x D x H)
* optionally available
Accessories
Electronic balance
Balance and elemental analyzer form a system. We offer analytical balances from the leading manufacturers with our instruments. Please ask us for further information.
Capsule sealing press
Tool for the exact dosing and gas-tight sealing of tin or silver capsules, for liquid and sensitive samples.
Sample former and pressing tools
For simplified sample preparation, we offer a manual tool for compressing samples in tinfoil or cups, and a mechanical press for the preparation of pellets with paper wrapping.
Coal standards
We offer a variety of coal standards with concentrations from < 0.5 % S up to 2.5 – 3.5 % S.
Unmatched analytical performance:
Limit of detection (LOD) of 10 ppm or < 15 μg C
- The soli TOC® cube is the most sensitive instrument in the market and offers an unmatched measurement range from 0.001 – 100 % carbon.
- Reproducible measurements due to precise adjustable temperature with fast heating rates. Temperature is measured where it matters, directly at the crucible.
- Consistent high measurement quality thanks to the use of a post-combustion catalyst. This ensures that even high carbon-content samples are quantitatively oxidized.
- Optimized separation of residual organic carbon (ROC) and total inorganic carbon (TIC) for special samples thanks to integrated carrier gas switching.
Fast & easy analyses: Autosampler for 24/7 operation & automated ash removal
- Autosampler with 89 positions, optimized for unattended 24/7 high-throughput operation.
- Minimized efforts for sample preparation due to the advanced temperature ramping and the crucible technology.
- Automated ash removal and clearly arranged, easily accessible system components minimize maintenance efforts.
Einfachste Probenvorbereitung: Minimierter Aufwand dank Tiegeltechnologie
- Einfache Analyse von Progeneinwaagen bis zu 3 g.
- Auch inhomogene Proben lassen sich zeitsparend einwiegen und präzise analysieren, da größere Probeneinwaagen den Aufwand für Homogenisierung deutlich reduzieren.
- Komfortable TOC-Messungen gemäß DIN EN 15936 mit direkter Ansäuerung der Probe im alternativen Keramiktiegel.
- Verringerte Kosten für Verbrauchsmaterialien durch wiederverwendbare Tiegel.

Unübertroffene Flexibilität: Anpassbar auf Ihre analytischen Anforderungen
- Optionale Bestimmung des Gesamtstickstoffs (TN) möglich.
- Die Analyse von Kohlenstoff in Flüssigkeiten ist für bestimmte Konzentrationen möglich.
- TOC400, ROC und TIC900 kann in nur einem Verfahrensschritt bestimmt werden. Hierzu werden die unterschiedlichen Bindungsformen des Kohlenstoffs mittels Temperaturrampe aus unterschiedlichen Zieltemperaturen unter oxidativen und pyrolytischen Bedingungen unterschieden.
- Alternativ kann auch das klassische Verfahren mittels Ansäuerung zur TIC Entfernung, Trocknung und anschließender TOC-Messung durchgeführt werden.
- DIN 19539 und DIN EN 15936/ISO 10694 konforme Analysen mit nur einem Gerät möglich.

Nutzerfreundliche Wartung: Komfortabel & werkzeuglos
- Wartung in Sekundenschnelle durch übersichtlich angeordnete, leicht zugängliche Systemkomponenten und ein werkzeugloses Klemmverbindungssystem.
- Die automatische Ascheentfernung sowie die herausragende Robustheit und Langlebigkeit des Gerätes reduzieren Wartungs- und Reparaturaufwände.
- Das werkzeuglose Klemmverbindungssystem sorgt jederzeit für ein dichtes, zuverlässiges Gerät ohne Leckage.